Agilent Technologies InfinityLab LC Series User Manual page 200

Isocratic and quaternary pumps
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

11
Hardware Information
Firmware Description
Firmware Updates
Firmware updates can be done with the Agilent Lab Advisor software with files
on the hard disk (latest version should be used).
Required tools, firmware and documentation are available from the Agilent web:
https://www.agilent.com/en-us/firmwareDownload?whid=69761
The file naming conventions are:
PPPP_RVVV_XXX.dlb, where
For instructions on firmware updates refer to section Replacing Firmware in
chapter "Maintenance" or use the documentation provided with the Firmware
Update Tools.
Update of main system can be done in the resident system only. Update of the
NOTE
resident system can be done in the main system only.
Main and resident firmware must be from the same set.
Resident System
Figure 20
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
PPPP is the product number, for example, 1315B for the G1315B DAD,
R the firmware revision, for example, A for G1315B or B for the G1315C DAD,
VVV is the revision number, for example 650 is revision 6.50,
XXX is the build number of the firmware.
Main FW update
Resident FW Update
Firmware Update Mechanism
Main System
200