Running The Test - Agilent Technologies InfinityLab LC Series User Manual

Isocratic and quaternary pumps
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

7
Test Functions and Calibration
System Pressure Test

Running the Test

Parts required
p/n
5067-6127
Damage to pressure sensitive parts
C AU T I O N
Even columns that are suitable for high pressures are sensitive to pressure
drops that occur during this test.
Do not include any pressure sensitive parts to the flow path and choose a
maximum pressure that is compatible to your system. For example, do not
include columns, a standard pressure flow cell (up to 20 bar) or a 400 bar
autosampler to a 600 bar pressure test.
Running the test from the Agilent Lab Advisor
1 Select the System Pressure Test from the Test Selection menu.
2 Start the test and follow the instructions.
Make sure to release the pressure by slowly opening the purge valve when the
NOTE
test has finished.
"Evaluating the Results"
of the System Pressure Test results. For detailed instructions refer to the Agilent
Lab Advisor software.
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
Description
A
Blank Nut SL
on page 116 describes the evaluation and interpretation
115