Maintenance

Maintenance of a Pump Head with Seal Wash Option

Maintenance of a Pump Head with Seal Wash
Option
For bio-inert modules use bio-inert parts only!
When
In case of maintenance or pump head internal leaks.
Tools required
Description
Wrench 1/4 inch
Hexagonal key, 4 mm
Parts required
#
1
1
OR
1
OR
1
1
1
OR
1
1
1
OR
1
1
Preparations
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
p/n
Description
A
01018-23702
Insert tool
A
0905-1175
Wash seal (PTFE)
A
0905-1718
Wash Seal PE
A
0905-1731
Bio-Inert Wash Seal
A
5062-2484
Gasket, seal wash (pack of 6)
A
5063-6589
Piston seal PTFE, carbon filled, black (pack of 2), default
A
G5611-21503
Piston Seal PTFE (Bio-inert)
A
0905-1420
PE seals (pack of 2)
A
5063-6586
Sapphire piston, 100 µL
A
5067-4695
Sapphire piston (Bio-inert)
A
8660-0852
Abrasive mesh
Switch off the pump at the main power switch.
Open the doors of the module.
Use an optional solvent shutoff valve or lift up solvent filters for avoiding leakages.
Remove the Pump Head Assembly.
Remove the wash solvent tubings from the support ring inlet and outlet.