Agilent Technologies InfinityLab LC Series User Manual page 230

Isocratic and quaternary pumps
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

12
LAN Configuration
Manual Configuration
5 Use the "/" and press Enter to list the current settings.
Figure 35
Telnet - Current settings in "Using Stored" mode
6 Change the IP address (in this example 192.168.254.12) and type "/" to list
Figure 36
Telnet - Change IP settings
7 When you have finished typing the configuration parameters, type
If the Initialization Mode Switch is changed now to "Using Stored" mode, the
NOTE
instrument will take the stored settings when the module is re-booted. In the
example above it would be 192.168.254.12.
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
current settings.
exit and press Enter to exit with storing parameters.
Figure 37
Closing the Telnet Session
information about the LAN interface
MAC address, initialization mode
Initialization mode is Using Stored
active TCP/IP settings
TCP/IP status - here ready
connected to PC with controller software (e.g. Agilent
ChemStation), here not connected
change of IP setting to
Initialization mode is Using Stored
active TCP/IP settings
stored TCP/IP settings in non-volatile memory
connected to PC with controller software (e.g. Agilent
ChemStation), here not connected
230