Exchange the Outlet Valve
1
Using a 1/4 inch wrench disconnect the valve capillary
from the outlet valve.
3
Reinstall the outlet valve and tighten it using a torque
wrench set to approx. 10 Nm.
5
Close the doors.
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
approx.
10 Nm
2
Using the 14 mm wrench loosen the valve and remove it
from the pump body.
NOTE
Do not disassemble the outlet valve, as this can damage the
valve.
4
Reconnect the valve capillary to the outlet valve.
135