Agilent Technologies InfinityLab LC Series User Manual page 156

Isocratic and quaternary pumps
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

8
Maintenance
Exchange the Multi-Channel Gradient Valve (MCGV)
1
Disconnect the solvent tubes from the MCGV.
3
Press the lower sides of the cover to unclip it. Remove
the cover.
5
Unscrew the two screws and remove them.
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
2
Disconnect the connecting tube from the MCGV.
4
Disconnect the MCGV cable from its connector.
6
Remove the MCGV.