2
Site Requirements and Specifications
Physical Specifications
Physical Specifications Agilent 1260 Infinity II
Quaternary Pump (G7111B)
Table 3
Type
Weight
Dimensions
(height × width × depth)
Line voltage
Line frequency
Power consumption
Ambient operating
temperature
Ambient non-operating
temperature
Humidity
Operating altitude
Safety standards:
IEC, EN, CSA, UL
ISM Classification
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
Physical Specifications G7111B
Specification
14.5 kg (32 lbs)
180 x 396 x 436 mm
(7.1 x 15.6 x 17.2 inches)
100 – 240 V~, ± 10 %
50 or 60 Hz, ± 5 %
80 VA, 65 W
4–55 °C (39–131 °F)
-40 – 70 °C (-40 – 158 °F)
< 95 % r.h. at 40 °C (104 °F)
Up to 3000 m (9842 ft)
Overvoltage category II,
Pollution degree 2
ISM Group 1 Class B
Comments
Wide-ranging capability
Non-condensing
For indoor use only
According to CISPR 11
37