Agilent Technologies InfinityLab LC Series User Manual page 144

Isocratic and quaternary pumps
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

8
Maintenance
Maintenance of a Pump Head Without Seal Wash Option
3
Remove the support rings from the piston housing.
5
Check the piston surface and remove any deposits or
layers: clean the piston surface with abrasive paper and
rinse with 2-propanol. Replace piston if scratched.
Piston Surface
7
Using the plastic side of the insert tool, insert new seals
into the pump head.
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
4
Lift the housing away from the pistons.
6
Using the steel side of the insert tool, carefully remove
the seal from the pump housing.
8
Place the support rings on the piston housing. Note the
correct position of the pins.
144