8
Maintenance

Replace Leak Handling System Parts

Replace Leak Handling System Parts

Parts required
p/n
5043-0856
5063-6527
1 Open the doors.
2 Press the Leak Adapter down and remove it together with the tubing.
3 Install the Leak Adapter by pressing it into the Main Cover.
4 Insert the Tubing (approximately 85 mm required for replacement) between
5 Close the doors.
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
Description
A
Leak Adapter
A
Tubing, Silicon Rubber, 1.2 m, ID/OD 6/9 mm
approximately 85 mm required
Leak Adapter outlet and Leak Panel.