Checking and Replacing the Solvent Filter

Checking and Replacing the Solvent Filter

A functional solvent filter is essential for a good pump performance and for
protecting the LC system.
When
If solvent filter is blocked.
Parts required
p/n
5041-2168
See
Small particles can permanently block the capillaries and valves of the module.
C AU T I O N
Damage of the module.
Always filter solvents.
Never use the module without solvent inlet filter.
If the filter is in good condition the solvent will freely drip out of the solvent tube
NOTE
(hydrostatic pressure). If the solvent filter is partly blocked only very little solvent
will drip out of the solvent tube.
1 Remove the solvent filter from the inlet filter adapter and replace it by a new
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
Description
A
Solvent inlet filter, 20 µm pore size
"Bottle Head Assembly"
one.
on page 182 for related parts.