Download Print this page
Agilent Technologies InfinityLab LC Series Installation Manual
Agilent Technologies InfinityLab LC Series Installation Manual

Agilent Technologies InfinityLab LC Series Installation Manual

Inlet/waste tubings replacement
Hide thumbs Also See for InfinityLab LC Series:

Advertisement

Quick Links

This technical note describes the installation of the inlet/waste tubing kit to a
1260 Infinity II Fraction Collector:
• Agilent InfinityLab LC Series 1260 Infinity II Preparative-Scale Fraction
Collector (G1364E)
• Agilent InfinityLab LC Series 1260 Infinity II Analytical-Scale Fraction
Collector (G1364F)
• Agilent InfinityLab LC Series 1260 Infinity II Bio-inert Fraction Collector
(G5664B)
Contents
Agilent InfinityLab LC Series
1260 Infinity II Fraction Collectors
Inlet / Waste Tubings Replacement
Replace the Inlet/Waste Tubings
Replace the Valve to Needle Tubing
2
8
Agilent Technologies

Advertisement

loading

Summary of Contents for Agilent Technologies InfinityLab LC Series

  • Page 1 • Agilent InfinityLab LC Series 1260 Infinity II Preparative-Scale Fraction Collector (G1364E) • Agilent InfinityLab LC Series 1260 Infinity II Analytical-Scale Fraction Collector (G1364F) • Agilent InfinityLab LC Series 1260 Infinity II Bio-inert Fraction Collector (G5664B) Contents Replace the Inlet/Waste Tubings...
  • Page 2: Replace The Inlet/Waste Tubings

    Replace the Inlet/Waste Tubings For bio-inert modules use bio-inert parts only! When • When contaminated, worn out or visibly damaged • Typically once every year Parts required Description Preparative-scale: G1364-68605 Tubing Kit 40 – 100 mL/min, 1.2 mm ID Standard tubing kit for Preparative Fraction Collector G1364-68603 Tubing Kit 4 –...
  • Page 3 Liquid spills or fraction losses C A U T I O N Worn or damaged tubings can cause potential spills or lead to fraction losses. ➔ Explicitly follow the described installation procedures to maximize the lifetime of inlet / waste tubing assembly and the valve to needle tubing, and to avoid potential spills or fraction losses.
  • Page 4 3 Unclip the inlet / waste tubing assembly from the leak adapter. 4 Remove the outer tubing kit from the unions. 5 Remove the leak adapter from the module housing. 6 Remove the unions from the inner tubing kit.
  • Page 5 7 Remove the clip by pushing it down through the module housing. 8 Remove the inner tubing kit by pulling it down through the opening in the module housing. 9 Slide in the new inner tubing kit from bottom to top (1.) and let the clip click into position (2.).
  • Page 6 11 Install the leak adapter to the module housing. 12 Connect the outer tubing kit to the unions. 13 Clip the inlet / waste tubing assembly into the leak adapter. 14 A color coded ring on one of the tubings and the valve body indicates, which tubing belongs to which port.
  • Page 7 15 Connect the finger-tight fittings of the inner tubing kit to the ports of the diverter valve. The tubings must run into the ports of the diverter valve in horizontal, parallel lines. N O T E The tubings must not be bent or twisted. 16 Connect the tubing with the label with the arrow (1) to the flow cell of the detector.
  • Page 8: Replace The Valve To Needle Tubing

    Replace the Valve to Needle Tubing For bio-inert modules use bio-inert parts only! When • When contaminated, worn out or visibly damaged • Typically once every year Tools required Description 8710-1534 Wrench, open end, 4 mm 8710-0510 Wrench open 1/4 — 5/16 inch Parts required Description Preparative-scale:...
  • Page 9 Liquid spills or fraction losses C A U T I O N Worn or damaged tubings can cause potential spills or lead to fraction losses. ➔ Explicitly follow the described installation procedures to maximize the lifetime of inlet / waste tubing assembly and the valve to needle tubing, and to avoid potential spills or fraction losses.
  • Page 10 3 Using the 4 mm wrench and the 5/16“ wrench for counter-holding, unscrew the valve to needle tubing from the needle. 4 Un-clip the tubing from the bottom of the needle carrier assembly. 5 Slide the tubing through the hole in the needle carrier assembly (from bottom to top) (1.) and out of the holder in the z-arm assembly (2.).
  • Page 11 6 Install the new valve to needle tubing assembly by clipping it in to the holder in the z-arm assembly (1.), and sliding it through the hole in the z-arm (2.) and out on the bottom of the needle carrier assembly (top to bottom). 7 Using the 4 mm wrench and the 5/16“...
  • Page 12 9 Screw the finger-tight fitting into the port of the diverter valve. 10 Re-install the tray(s) in the tray base. 11 Start the instrument. 12 Close the front cover. © *G1364-90200* Part Number: Agilent Technologies, Inc 2017 G1364-90200 *G1364-90200* Agilent Technologies, Inc G1364-90200 Edition: 08/2017 Hewlett-Packard-Strasse 8...

This manual is also suitable for:

1260 infinity ii