General Error Messages

General Error Messages

General error messages are generic to all Agilent series HPLC modules and may
show up on other modules as well.

Timeout

Error ID: 0062
The timeout threshold was exceeded.
Probable cause
1
2
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
The analysis was completed successfully,
and the timeout function switched off the
module as requested.
A not-ready condition was present during a
sequence or multiple-injection run for a
period longer than the timeout threshold.
Suggested actions
Check the logbook for the occurrence and source
of a not-ready condition. Restart the analysis
where required.
Check the logbook for the occurrence and source
of a not-ready condition. Restart the analysis
where required.