8
Maintenance
Exchange the Active Inlet Valve (AIV) or its Cartridge
Exchange the Active Inlet Valve (AIV) or its
Cartridge
For bio-inert modules use bio-inert parts only!
When
If internally leaking (backflow)
Tools required
Description
Wrench, 14 mm
Pair of tweezers
Parts required
p/n
G5699A
G1312-60025
G1312-60020
G1311-67304
G1311-61601
0100-2298
G5611-60025
G5611-60020
Preparations
•
•
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
Description
A
Active Inlet Valve Upgrade Kit
A
Active inlet valve body, without cartridge
A
Cartridge for active inlet valve 600 bar
A
Connecting tube
A
Active Inlet Valve Cable Assembly
A
Adapter, PEEK int. 1/4-28 to ext. 10-32
A
Active Inlet Valve (AIV, Bio-inert), without cartridge
A
Cartridge for AIV (Bio-inert)
Switch off pump at the main power switch and unplug the power cable.
Use an optional solvent shutoff valve or lift up solvent filters in solvent reservoirs for avoiding
leakages.
160