Module Error Messages; Solvent Zero Counter; Pressure Above Upper Limit - Agilent Technologies InfinityLab LC Series User Manual

Isocratic and quaternary pumps
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

6
Error Information

Module Error Messages

Module Error Messages

These errors are pump specific.

Solvent Zero Counter

Error ID: 2055, 2524
Pump firmware version A.02.32 and higher allow to set solvent bottle fillings in
the data system. If the volume level in the bottle falls below the specified value
the error message appears when the feature is configured accordingly.
Probable cause
1
2

Pressure Above Upper Limit

Error ID: 2014, 2500
The system pressure has exceeded the upper pressure limit.
Probable cause
1
2
3
4
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
Volume in bottle below specified volume.
Incorrect setting.
Upper pressure limit set too low.
Blockage in the flowpath (after the damper).
Defective damper.
Defective mainboard.
Suggested actions
Refill bottles and reset solvent counters.
Make sure the limits are set correctly.
Suggested actions
Ensure the upper pressure limit is set to a value
suitable for the analysis.
Check for blockage in the flow path. The
following components are particularly subject to
blockage: inline filter frit, needle (autosampler),
seat capillary (autosampler), sample loop
(autosampler), column frits and capillaries with
small internal diameters (e.g. 50 µm ID).
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.