Agilent Technologies InfinityLab LC Series User Manual page 33

Isocratic and quaternary pumps
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

2
Site Requirements and Specifications
Site Requirements
Power Cords
Country-specific power cords are available for the module. The female end of all
power cords is identical. It plugs into the power-input socket at the rear. The male
end of each power cord is different and designed to match the wall socket of a
particular country or region.
Agilent makes sure that your instrument is shipped with the power cord that is
suitable for your particular country or region.
Unintended use of power cords
WAR N IN G
Using power cords for unintended purposes can lead to personal injury or
damage of electronic equipment.
Never use a power cord other than the one that Agilent shipped with this
instrument.
Never use the power cords that Agilent Technologies supplies with this
instrument for any other equipment.
Never use cables other than the ones supplied by Agilent Technologies
to ensure proper functionality and compliance with safety or EMC
regulations.
Absence of ground connection
WAR N IN G
The absence of ground connection can lead to electric shock or short
circuit.
Never operate your instrumentation from a power outlet that has no
ground connection.
Electrical shock hazard
WAR N IN G
Solvents may damage electrical cables.
Prevent electrical cables from getting in contact with solvents.
Exchange electrical cables after contact with solvents.
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual