Hardware Information

Electrical Connections

Electrical Connections

Electrical Connections
Never use cables other than the ones supplied by Agilent Technologies to ensure
NOTE
proper functionality and compliance with safety or EMC regulations.
Electric Shock
Electric shock due to insufficient insulation of connected instruments
WAR N IN G
Personal injury or damage to the instrument
Any other instruments connected to this instrument shall be approved to
a suitable safety standard and must include reinforced insulation from
the mains.
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
The CAN bus is a serial bus with high-speed data transfer. The two
connectors for the CAN bus are used for internal module data transfer and
synchronization.
The ERI/REMOTE connector may be used in combination with other analytical
instruments from Agilent Technologies if you want to use features such as
start, stop, common shutdown, prepare, and so on.
With the appropriate software, the LAN connector may be used to control the
module from a computer through a LAN connection. This connector is
activated and can be configured with the configuration switch.
With the appropriate software, the USB connector may be used to control the
module from a computer through a USB connection.
The power input socket accepts a line voltage of 100 – 240 VAC ± 10 % with a
line frequency of 50 or 60 Hz. Maximum power consumption varies by
module. There is no voltage selector on your module because the power
supply has wide-ranging capability. There are no externally accessible fuses
because automatic electronic fuses are implemented in the power supply.