8
Maintenance
Reinstall the Pump Head Assembly
3
Insert the pump head screws and tighten them
stepwise. First tighten the left screw, then the right
screw, with a torque of approximately 5 Nm.
approx.
5 Nm
5
Reconnect the valve capillary to the outlet valve.
7
Reconnect the outlet capillary and the waste tubing.
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
4
Install the capillary at the bottom of the pump head.
6
Reconnect the solvent inlet tube to the passive inlet
valve.
8
If installed, reconnect the active inlet valve cable to its
connector.