Maintenance

Exchange the Outlet Valve

Exchange the Outlet Valve

For bio-inert modules use bio-inert parts only!
When
If internally leaking
Tools required
p/n
8710-0510
8710-1924
5067-5688
Parts required
#
1
OR
1
OR
1
Preparations
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
Description
A
Open-end wrench 1/4 — 5/16 inch
A
Open-end wrench 14 mm
A
Torque wrench 1 – 25 Nm with 14 mm wrench
p/n
Description
A
G1312-60067
Outlet valve
A
G1312-60167
Outlet valve (type N/SFC)
A
G5611-60067
Outlet Valve (Bio-Inert)
Switch off the pump at the main power switch.
Open the doors of the module.