Remote Cables - Agilent Technologies 1260 Infinity II User Manual

Binary pump
Hide thumbs Also See for 1260 Infinity II:
Table of Contents

Advertisement

10
Identifying Cables

Remote Cables

Remote Cables

ERI (Enhanced Remote Interface)
• 5188-8029 ERI to general purpose (D-Sub 15 pin male - open end)
• 5188-8044 ERI to ERI (D_Sub 15 pin male - male)
• 5188-8059 ERI-Extension-Cable 1.2 m (D-Sub15 pin male / female)
p/n 5188-8029
176
pin
Color code
Enhanced
Remote
1
white
IO1
2
brown
IO2
3
green
IO3
4
yellow
IO4
5
grey
IO5
6
pink
IO6
7
blue
IO7
8
red
IO8
9
black
1wire DATA
10
violet
DGND
11
grey-pink
+5V ERI out
12
red-blue
PGND
13
white-green
PGND
14
brown-green
+24V ERI out
15
white-yellow
+24V ERI out
NC
yellow-brown
InfinityLab LC Series 1260 Infinity II Binary Pump User Manual
Classic
Active (TTL)
Remote
START
Low
REQUEST
STOP
Low
READY
High
POWER ON
High
NOT USED
SHUT DOWN
Low
START
Low
PREPARE
Low