7
Test Functions and Calibration
Leak Rate Test

Evaluating the Results

Results of the leak rate test are the leak rates measured for pistons 1 and 2 as
described for the test principle. If any of the leak rates exceeds 3 µL/min, the test
will fail.
Inlet valve
Pump chamber 1
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
Damper
Outlet valve
Seals
Piston 1
to column
Purge valve
Pump chamber 2
Piston 2
to waste