LAN Configuration
Manual Configuration
3 Type
Table 30
Value
?
/
ip <x.x.x.x>
sm <x.x.x.x>
gw <x.x.x.x>
exit
4 To change a parameter follows the style:
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
? and press enter to see the available commands.
Figure 34
Telnet Commands
Telnet Commands
•
parameter value, for example:
ip 134.40.28.56
Then press [Enter], where parameter refers to the configuration parameter
you are defining, and value refers to the definitions you are assigning to that
parameter. Each parameter entry is followed by a carriage return.
Description
displays syntax and descriptions of commands
displays current LAN settings
sets new ip address
sets new subnet mask
sets new default gateway
exits shell and saves all changes