Replace Leak Handling System Parts

Replace Leak Handling System Parts

Parts required
p/n
5063-6527
1
Open the doors.
5
Close the doors.
118
Description
Tubing assembly, i.d. 6 mm, o.d. 9 mm, 1.2 m (to waste)
approximately 85 mm required
2
Press the Leak Adapter down and remove it together
with the tubing.
3
Install the Leak Adapter by pressing it into the Main
Cover.
4
Insert the Tubing (approximately 85 mm required for
replacement) between Leak Adapter outlet and Leak
Panel.
1260 SFC Binary Pump