Replace Leak Handling System Parts - Agilent Technologies 1260 Infinity II User Manual

Binary pump
Hide thumbs Also See for 1260 Infinity II:
Table of Contents

Advertisement

8
Maintenance

Replace Leak Handling System Parts

Replace Leak Handling System Parts

Parts required
p/n
5063-6527
1
Open the doors.
3
Install the Leak Adapter by pressing it into the Main
Cover.
142
Description
Tubing assembly, i.d. 6 mm, o.d. 9 mm, 1.2 m (to waste)
approximately 85 mm required
2
Press the Leak Adapter down (1.) and remove it together
with the tubing (2.).
4
Insert the Tubing (approximately 85 mm required for
replacement) between Leak Adapter outlet and Leak
Panel.
InfinityLab LC Series 1260 Infinity II Binary Pump User Manual