8
Maintenance

Exchange the Outlet Valve

Exchange the Outlet Valve

When
if leaking internally
Tools required
p/n
8710-1924
8710-0510
5067-5688
Parts required
#
1
2
Preparations
Switch off the pump at the main power switch
1
Using a ¼ inch wrench disconnect the absorber capillary
from the outlet valve.
144
Description
Open-end wrench 14 mm
Open-end wrench 1/4 — 5/16 inch
Torque wrench 1 – 25 Nm with 14 mm wrench
p/n
Description
G1312-60167
Outlet valve Channel A
G1312-60067
Outlet valve Channel B
InfinityLab LC Series 1260 Infinity II Binary Pump User Manual
2
Unscrew the valve with the 14 mm wrench and remove it
from the pump body.
N O T E
Do not disassemble the outlet valve, as this can
damage the valve.