8
Maintenance

Exchange the Outlet Valve

Exchange the Outlet Valve

When
if leaking internally
Tools required
Description
Wrench, 1/4 - 1/5 inch
Wrench 1/4 inch
Wrench, 14 mm
Parts required
#
1
2
Preparations
Switch off the pump at the main power switch
1
Using a ¼ inch wrench disconnect the absorber capillary
from the outlet valve.
120
p/n
Description
G1312-60167
Outlet valve Channel A
G1312-60067
Outlet valve Channel B
2
Unscrew the valve with the 14 mm wrench and remove it
from the pump body.
N O T E
Do not disassemble the outlet valve, as this can
damage the valve.
1260 SFC Binary Pump