Adjusting The 20° Position - Siemens MODULARIS Uro Plus Service Instructions Manual

Hide thumbs Also See for MODULARIS Uro Plus:
Table of Contents

Advertisement

Iso-center with X-Ray
5
6
Fig. 4
C
Fig. 6
Adjusting the 20° position
NOTICE
Move the C-arm into one of the two 20° positions.
Switch FL ON briefly.
The ball must be in the center of the cross (refer to Fig. 1).
Move the C-arm into the other 20° position.
Switch FL ON briefly.
The ball on the adjustment area must be located in the center of the centering cross
(refer to Fig. 1).
The distance from the phantom to the center of the centering cross must be the same for
both 20° positions (with a maximum deviation of ± 2 mm) (refer to C/Fig. 6).
Siemens AG
Medical Solutions
7
D
with 6/Fig.4
The position of the ball must not be identical in both 20° positi-
ons. The tolerances must however be maintained, refer to
Figure 1.
SPL1-130.840.02
Rev. 03
04.05
Fig. 5
E
RBV
RBV
with 6/Fig.4
Page 3 of 8
CS PS 24
5 - 3
8
5
MODULARIS Uro Plus