Siemens MODULARIS Uro Plus Service Instructions Manual page 67

Hide thumbs Also See for MODULARIS Uro Plus:
Table of Contents

Advertisement

LITHOSTAR MODULARIS
12
Fig. 32
From support arm serial number 0051
- Remove all cover screws.
Remove upper portion of covers (11/Fig. 30).
Couple the LITHOSTAR MODULARIS to the SIREMOBIL Iso-C.
Raise the vertical column of the SIREMOBIL Iso-C until the LITHOSTAR MODULARIS
hooks onto the docking plate of the SIREMOBIL Iso-C (12/Fig. 32).
Loosen the lock nut (13/Fig. 33).
Rotate the threaded nut (14/Fig. 33) to adjust the cam (15/Fig. 33) so that the switch is
rotated. Upward movement of the SIREMOBIL Iso-C is blocked.
Tighten the lock nut (13/Fig. 33).
Temporarily attach the braking handle.
Lock the LITHOSTAR MODULARIS and the SIREMOBIL Iso-C together.
Downward movement must be blocked.
If not:
- loosen the set screw (16/Fig. 33).
- rotate the lever until the switch is activated.
- tighten the set screw (16/Fig. 33).
Reattach the covers.
Siemens AG
Medical Solutions
17
Fig. 33
SPL1-130.840.02
Page 27 of 34
Rev. 03
04.05
CS PS 24
4 - 27
17
16
13
14
15
MODULARIS Uro Plus