Adjusting The 0° Position - Siemens MODULARIS Uro Plus Service Instructions Manual

Hide thumbs Also See for MODULARIS Uro Plus:
Table of Contents

Advertisement

5 - 6
Fig. 7
1
Fig. 9
Adjusting the 0° position
Couple the LITHOSTAR MODULARIS to the SIREMOBIL Iso-C as described in the
operating instructions. Do not couple the motor yet.
Manually perform the orbital movement (angulation range approx. ± 30°).
Set the SIREMOBIL Iso-C to orbital and angulation direction 0°.
Couple the motor.
Flip up the iso-center phantom.
Turn fluoro ON briefly. The ball must now be in the center of the cross. (refer to Fig. 8).
If not:
- Slightly loosen the three screws on the RBV (1/Fig. 9).
- Adjust the component with the crossing hair so that the ball is located in the cross hairs
(refer to Fig. 8).
- After adjusting the setting, retighten all screws.
If adjustment is not possible, adjust as for the support arm - serien no. < 0050.
MODULARIS Uro Plus
Iso-center with X-Ray
ok
requires
correction
Fig. 8
SPL1-130.840.02
Page 6 of 8
Rev. 03
04.05
CS PS 24
acceptable
Siemens AG
Medical Solutions
5