LITHOSTAR MODULARIS

Curved Probe

Loosen the 8 screws (4 are not visible) (8/Fig. 40) and remove the sleeve (7/Fig. 40).
Attach the sleeve on the new probe. Pay particular attention to the positioning, the
narrow side with the orange dot (6/Fig. 40) and (5/Fig. 40) must form a line.
Tighten the 8 screws.
Check the ultrasound iso-center.

Concluding work

Perform the protective conductor measurement.
Test all functions from the control panel.