LITHOSTAR MODULARIS Error List
Monitoring, E31
E31
Error Text:
Possible
cause/action:
E32
Error Text:
Possible
cause/action:
E34
Error Text:
Possible
cause/action:
Siemens AG
Medical Solutions
E34
Temperature sensor supplies incorrect data.
Temperature sensor o.k.?
Connection cable D3.X2 o.k.?
The temperature difference can be read out on the service PC.
The temperature at the shock wave head has exceeded the
maximum value.
Let the shock wave head cool down, but do not switch the
system off!
Check that the cooling unit is operating.
If M1 cooling pump is running pump head may need to be repla-
ced.
If the error recurs, check the function of the cooling unit.
Temperature cannot be determined: could not access the ADC
when converting the analog values from the temperature sen-
sor.
Voltage supply for the temperature sensors outside acceptable
range.
Connection cable D3.X2 o.k.?
Board D3?
Read out the temperature with the service PC.
Refer also to error 45.