Siemens MODULARIS Uro Plus Service Instructions Manual page 87

Hide thumbs Also See for MODULARIS Uro Plus:
Table of Contents

Advertisement

Iso-center with ultrasound
Fig. 9
Distance check: height
Press the "Measurement" key.
With the trackball set the cross on center of target cross.
Press the Measurement key.
Move the trackball until the cross lies in the center of the isocenter cross (1/Fig. 9).
Press the "Measurement" key.
Move the cross "+2" with the trackball on the longitudinal axis of the target line into the
isocenter.
Press the
Move the cross with the trackball horizontally into the center of the white area on the
isocenter phantom.
The displayed value for "D2" (2/Fig. 9) must be between 1 and 5 mm. This has validity
only for the default values, see instructions SPL1-130.038.01.. .
Press the
Distance check: side
Press the
Move the cross "+2" with the trackball onto the longitudinal axis of the target line, at the
height of the isocenter.
Press the
Move the cross with the trackball horizontally into the center of the white area on the
isocenter phantom.
Siemens AG
Medical Solutions
1
2
key.
key.
key two times (cross "+1" is not required in this measurement).
key.
SPL1-130.840.02
Rev. 03
Page 5 of 16
04.05
CS PS 24
6 - 5