High voltage circuit, E60
LITHOSTAR MODULARIS Circuit diagram J1048 - 12.
E60
Error Text:
Possible
cause/action:
E61
Error Text:
Possible
cause/action:
E62
Error Text:
Possible
cause/action:
E63
Error Text:
Possible
cause/action:
E64
Error Text:
Possible
cause/action:
E65
Error Text:
Possible
cause/action:
E66
Error Text:
Possible
cause/action:
E67
Error Text:
Possible
cause/action:
MODULARIS Uro Plus
LITHOSTAR MODULARIS Error List
E67
Charging current too low during the charging process.
Check the line voltage!
Replace the charging / energy unit!
Overvoltage in the charging unit.
System OFF
- 10 minute pause -
System ON.
Replace the charging / energy unit!
Overtemperature in the charging unit.
Heat in the LITHO MODULARIS ?
System OFF
- 10 minute pause -
System ON.
Replace the charging / energy unit!
Capacitor defective in the charging / energy unit (out of tole-
rance).
Replace the charging / energy unit!
Default voltage not attained.
Current voltage after frequency converter (U
selected voltage.
Check the fiber optic cable!
Replace the charging / energy unit!
Charging voltage too high during the charging process.
Replace the charging / energy unit!
Default voltage incorrect.
Check the fiber optic cable!
Replace the charging / energy unit!
No ignition of spark gap after trigger pulse (several times).
Check the fiber optic cable!
Replace the charging/ energy unit!
SPL1-130.840.02
Rev. 03
04.05
) is greater than
nom
1
)
1
)
1
)
Page 8 of 14