2 - 10
High voltage circuit, E75
E75
Error Text:
Possible
cause/action:
E76
Error Text:
Possible
cause/action:
E77
Error Text:
Possible
cause/action:
E78
Error Text:
Possible
cause/action:
E79
Error Text:
Possible
cause/action:
1
) Before replacing the charging / energy unit, check the fiber optic cable (LWL):
- Disconnect the fiber optic cable on both ends and check by holding it up to the light
- place it on the Clk (U5) of board D3 and check whether light is visible.
MODULARIS Uro Plus
LITHOSTAR MODULARIS Error List
–
E78
Chip card terminal error (documented in error log only).
Check execution level for board D3:
AS 00 - 02: replace board D3.
AS 03 and higher: replace chip card reading device.
Wrong card or card incorrectly inserted.
Chip card expired
Insert a new card.
Chip card communication error
Check execution level for board D3:
AS 00 - 02: replace board D3.
AS 03 and higher: replace chip card reading device
Cable?
If the error recurs after replacing board D3, is plug X33 properly
connected?
Charging process too long.
Line voltage o.k.?
Replace the charging / energy unit!
Call Service
HV-cable ok?
Short circuit?
or
SPL1-130.840.02
Rev. 03
Page 10 of 14
04.05
CS PS 24
2
Siemens AG
Medical Solutions