2 - 10
High voltage circuit, E75
E75
Error Text:
Possible
cause/action:
E76
Error Text:
Possible
cause/action:
E77
Error Text:
Possible
cause/action:
E78
Error Text:
Possible
cause/action:
E79
Error Text:
Possible
cause/action:
1
) Before replacing the charging / energy unit, check the fiber optic cable (LWL):
- Disconnect the fiber optic cable on both ends and check by holding it up to the light
- place it on the Clk (U5) of board D3 and check whether light is visible.
MODULARIS Uro Plus
LITHOSTAR MODULARIS Error List
E78
Chip card terminal error (documented in error log only).
Check execution level for board D3:
AS 00 - 02: replace board D3.
AS 03 and higher: replace chip card reading device.
Wrong card or card incorrectly inserted.
Chip card expired
Insert a new card.
Chip card communication error
Check execution level for board D3:
AS 00 - 02: replace board D3.
AS 03 and higher: replace chip card reading device
Cable?
If the error recurs after replacing board D3, is plug X33 properly
connected?
Charging process too long.
Line voltage o.k.?
Replace the charging / energy unit!
Call Service
HV-cable ok?
Short circuit?
or
SPL1-130.840.02
Rev. 03
Page 10 of 14
04.05
CS PS 24
2
Siemens AG
Medical Solutions