4 - 28
Fig. 34
24 -1 mm
Fig. 36
Balancing Spring for the Support Arm
•
Position the support arm vertically.
•
Remove covers from the cart.
•
Remove the cover from the column/mount for the ECG triggering.
•
Remove the cover from the high voltage connector on the charge device (this simplifies
removing the spring mount).
•
Remove the 3 screws (1/Fig. 34). Take out the spring mount.
•
Remove the spring mount.
•
Replace the balancing spring.
- If the mount is different from the ones delivered, it must be replaced as well.
- Be sure to have correct spacing (Fig. 36).
•
Reinstall the spring mount.
MODULARIS Uro Plus
1
1
SPL1-130.840.02
Rev. 03
LITHOSTAR MODULARIS
Fig. 35
Page 28 of 34
04.05
CS PS 24
1
ok
ok
not ok
4
Siemens AG
Medical Solutions