Air Suction Hose; To Support Arm Serial Number 0050 - Siemens MODULARIS Uro Plus Service Instructions Manual

Hide thumbs Also See for MODULARIS Uro Plus:
Table of Contents

Advertisement

4 - 18
2
Fig. 23

Air suction hose

To support arm serial number 0050

Remove the covers of the basic system unit and the shock wave head.
Remove the shock wave head (refer to the corresponding section).
Pull the temperature sensor (T/Fig. 18) in the shock wave head support somewhat
forward and then push it back through to the side.
Secure the spiral spring to the sensor supply cable (x/Fig. 19) approx. 4 cm behind the
temperature sensor with adhesive tape (KB/Fig. 19) to keep it from slipping back.
Remove the white teflon plate (Tp/Fig. 18) by removing the screws.
Push the high-voltage connector (W/Fig. 20) slightly to the side for better access to the
water supply (Fig. 20).
Now pull the water supply (W/Fig. 20) with the center air suction hose slightly forward in
the shock wave head support and then push it back through to the side.
Remove the front bracket connection (F/Fig. 25) on the water supply by removing the
screws.
Remove the front part (E/Fig. 25) of the water supply by removing the screws.
Disconnect the old water hose.
Install the new water hose and secure it with the appropriate hardware.
Remove the front part (E/Fig. 25) of the water supply by removing the screws.
Puncture the tubing with a needle (app. 1 mm) twice at this point (<-/Fig. 25).
Insulate the bracket connection of the water supply (F/Fig. 25) just removed, i.e. attach
new teflon insulation tape with the screws; when doing this do not block the opening to
the pressure sensor.
MODULARIS Uro Plus
LITHOSTAR MODULARIS
5
4
Fig. 24
SPL1-130.840.02
Page 18 of 34
Rev. 03
04.05
CS PS 24
6
3
4
4
Siemens AG
Medical Solutions