LITHOSTAR MODULARIS Error List
High voltage circuit, E68
E68
Error Text:
Possible
cause/action:
E69
Error Text:
Possible
cause/action:
E70
Error Text:
Possible
cause/action:
E71
Error Text:
Possible
cause/action:
E72
Error Text:
Possible
cause/action:
E73
Error Text:
Possible
cause/action:
E74
Error Text:
Possible
cause/action:
1
) Before replacing the charging / energy unit, check the fiber optic cable (LWL):
- Disconnect the fiber optic cable on both ends and check by holding it up to the light
- place it on the Clk (U5) of board D3 and check whether light is visible.
Siemens AG
Medical Solutions
E74
Charging current too high during charging process.
Replace the charging / energy unit!
Repeated self-ignition of the spark gap.
Replace the charging/ energy unit!
Charging current too low after ignition!
Check the connections of the high voltage cable!
Check the high voltage connector!
Replace the shock wave head!
Charging current too high after ignition.
Check the high voltage cable!
Check the high voltage connector!
Replace the shock wave head!
No communication possible with the charging / energy unit.
Check the line voltage at the charging / energy unit!
Replace the fuse in the charging / energy unit!
Check the fiber optic cable!
Replace the charging / energy unit!
Message: Constancy limit reached.
For user information only.
Approx. 10-15 treatments still possible, then error 74 appears.
Constancy limit reached.
The constancy limit has been reached.
Replace the following parts:
- shock wave head
In addition, replace the water in the entire system!