Water system, E14
E14
Error Text:
Possible
cause/action:
E15
Error Text:
Possible
cause/action:
Controls, E20
If the control panel is dark: check fuse F5 on board D3.
E20
Error Text:
Possible
cause/action:
E21
Error Text:
Possible
cause/action:
MODULARIS Uro Plus
LITHOSTAR MODULARIS Error List
E15
Water level too low in the water reserve tank.
Not enough water in the tank.
Float switch in the tank o.k.?
Board D3.X3 connection o.k.?
Completely fill the coupling bellows, deselect automatic pres-
sure regulation, and drain the water.
Repair any leaks.
Refill the water in the reserve tank.
Coupling pressure not o.k.
Pressure sensor on the shock head o.k.?
If water can still be pumped out, the sensor is probably defec-
tive (refer to "Working with the terminal program: show diagno-
stic data").
Suction hose correctly inserted in the coupling bellows?
Return hose kinked?
Board D3.X8 connection o.k.?
Fill the coupling circuit.
If the error only occurs with obese patients, the suction hose in
the coupling bellows may be kinked.
E21
No contact with the control unit (this error is visible only when
reading out the error log with the service PC).
Connection cable X5 correctly connected to the control unit?
Check fuse F5 on board D3.
If the error occurs after replacing board D3 - is plug X35 pro-
perly connected?
X-ray button on the control unit activated with switch on.
X-ray button stuck?
Control unit o.k.?
Cable o.k.?