Iso-center with ultrasound
D
Fig. 14
Make the following basic settings at the ultrasound unit:
Value:
- 6 dB (C/Fig. 14)
20 dB (B/Fig. 14)
3 (A/Fig. 14)
60 (D/Fig. 14)
60 dB (E/Fig. 14)
Siemens AG
Medical Solutions
E
A
B
C
Adjust using:
Set trackball to SI
(white background),
then actuate
set value with control
Knob B
Set trackball to FV
(white background),
then actuate
set value with control
Knob
Set trackball to DR
(white background),
then actuate
set value with control
SPL1-130.840.02
Rev. 03
Fig. 15
German
English
and
SI
Output
EV
Gain
and
FV
Edge
-------
-------
and
DR
Dyn Rng
Page 9 of 16
04.05
CS PS 24
6 - 9
1
French
Spanish
Sortie
Salida
Gain
Ampl
Côté
Contorno
-------
-------
DR
Cam Din