CONVEX APPLICATOR
-----------------------------------------
actual values
-----------------------------------------
threshold value =
scale low
=
scale high
=
Every probe can be set to one of three settings:
threshold value
fig. 2
scale low
fig. 2
scale high
fig. 2
By selecting the keys "t, l, h " (depending on application) the value is reduced; select T, L,
H to increase the value.

select option: r

Press the "r" key to change the "pulse rate"
Selection: fixed 2 Hz (USA) or variable 1/1.5 /2 Hz
After pressing the "h" key, display for "selection option r" shows the pulse rate set.

select option: s

CAUTION
Software downloads are possible only if the software and hardware are compatible.
Press the "s" key to display the following text:
=> S
press key '1' to confirm download, any other key to abort
After "1" has been entered, the software download must start within 60 seconds.
If the 60 seconds elapses without starting the download, reselect "h" from the entry menu,
reselect "s", etc.
Once the software download is complete, check that the values in the "ultrasound localiza-
tion" menu for the applicators are identical to those in document SPL1-130.038.01..
(Intranet). If this is not the case, make the appropriate changes.
MODULARIS Uro Plus
change with key
(-) (+)
100 mm
t
7.7 %
l
7.7 %
h
The point at which the "scale high" value is used (default 100 mm).
Correction factor (default 7.7 %) between 0 mm and the "threshold
value " (This value is useful for thin patients).
Correction factor (default 7.7 %) between "threshold value " and end
(This value is useful for obese patients.).
Software download.
If not observed, D3 board can be damaged.
The system and service PC must not be turned off during a soft-
ware download. If this occurs, board D3 must be replaced.
SPL1-130.840.02
Rev. 03
T
L
H
Page 12 of 18