Siemens MODULARIS Uro Plus Service Instructions Manual page 33

Hide thumbs Also See for MODULARIS Uro Plus:
Table of Contents

Advertisement

Service software
Convex
Array
Fig. 1
CAUTION
change values for sector applicator - Press key 's'
change values for convex applicator - Press key 'c'
Explanation:
The displayed position (distance and position) of the objects on the ultrasound image is
calculated from the time difference between sending a signal and receiving the echo.
The standard speed of sound in the human body is 1540 m/sec, however, this speed can
vary depending on which tissue is being examined.
The position of the iso-center can be adjusted to ensure that the position of the cross mat-
ches the actual stone position on the ultrasound monitor.
This is possible with:
SECTOR APPLICATOR
-----------------------------------------
actual values
-----------------------------------------
threshold value =
scale low
=
scale high
=
Siemens AG
Medical Solutions
Sector
The following values can only be altered if approval is given by
Erlangen.
The correct values can be found in document SPL1-130.038.01..
(Intranet).
change with key
(-) (+)
100 mm
t
7.7 %
l
7.7 %
h
SPL1-130.840.02
Rev. 03
Fig. 2
T
L
H
Page 11 of 18
04.05
CS PS 24
3 - 11
MODULARIS Uro Plus