Siemens MODULARIS Uro Plus Service Instructions Manual page 56

Hide thumbs Also See for MODULARIS Uro Plus:
Table of Contents

Advertisement

4 - 16
Tp
Fig. 18
Cables in the corrugated tubing
NOTICE
Remove the covers on the basic system unit and the shock wave head.
Remove the shock wave head (refer to the corresponding section).
Push the temperature sensor slightly forward (T/Fig. 18) in the shock wave head support
and then push it back to the side.
Secure the spiral spring to the sensor supply cable approx. 4 cm (x/Fig. 19) behind the
temperature sensor with adhesive tape to keep it from slipping back (KB/Fig. 19).
To support arm serial number 0050
- Remove the white teflon plate (Tp/Fig. 18) by removing the screws.
- Now pull the water supply (W/Fig. 20) in the shock wave head somewhat forward and
then push it back to the side.
From support arm serial number 0051
- Remove the two screws (1/Fig. 21).
- Remove the lower part (2/Fig. 21).
Remove the column cover (Fig. 23).
Do not cut open the cable that fastens the corrugated tubing to the support arm. It can be
opened with a small movement of the screw.
Unscrew the corrugated tubing (2/Fig. 23).
Remove the disk (3/Fig. 24) from the corrugated tubing.
Replace the corresponding cable in the corrugated tubing.
- If the "temperature sensor" cable is being replaced, the water supply from the defective
one must be transferred to the new pressure sensor (for more information on this
procedure, refer to the section "Air suction hose").
MODULARIS Uro Plus
T
Fig. 19
All cables must be disconnected from the shock wave head in
order to replace parts of the corrugated tubing.
SPL1-130.840.02
Rev. 03
LITHOSTAR MODULARIS
T
x
KB
Fig. 20
Page 16 of 34
04.05
CS PS 24
W
T
Siemens AG
Medical Solutions