LITHOSTAR MODULARIS

Emptying the cooling circuit with the new pump (3/Fig. 11)

Remove the system covers from the basic unit.
Remove the water reserve tank and empty the water.
Insert the return hose into the water reserve tank.
Position the support arm in the horizontal position.
Open valve Y8 (Fig. 11),
i.e. switch the Y8 lever from horizontal = closed to vertical = open.
Exchange the two water connections on the cooling pump. (The two black-marked hoses
are not connected).
Switch the system on.
Set service switch S2 on board D3 to position 2 (service on).
Select "service" and then "cooling unit" on the control unit.
Activate the corresponding key for "empty" on the control unit until the water stops.
Set service switch S2 on board D3 to position 2 (service off).
Switch the system off.
Properly connect the two water connections to the cooling pump. (Both black-marked
hoses must be reconnected).

Emptying the coupling circuit

Remove the system covers on the basic unit.
Remove the water reserve tank and empty the water.
Switch the system on.
Position the support arm in the horizontal position.
Set service switch S2 on board D3 to position 2 (service on).
Select "service" and then "coupling circuit" on the control unit.
Activate the corresponding key for "empty" (S6/Fig. 4/ page 3-16) on the control unit until
the coupling bellows is located next to the lens.
Switch the system off.