2 - 6
Monitoring, E40
E40
Error Text:
Possible
cause/action:
E41
Error Text:
Possible
cause/action:
E42
Error Text:
Possible
cause/action:
E43
Error Text:
Possible
cause/action:
E44
Error Text:
Possible
cause/action:
E45
Error Text:
Possible
cause/action:
E46
Error Text:
Possible
cause/action:
E47
Error Text:
Possible
cause/action:
E48
Error Text:
Possible
cause/action:
MODULARIS Uro Plus
LITHOSTAR MODULARIS Error List
E48
Cooling pump fuse defective.
Fuse F2 on board D3 o.k.?
Cooling pump M1 defective.
Motor M1 o.k.?
Circuit diagram J1048 – 13
Connection cable D3.X7 o.k.?
Coupling pump defective.
Motor M2 o.k.?
Circuit diagram J1048 – 13
Connection cable D3.X7 o.k.?
Fuse for coupling pump defective.
Fuse F3 on board D3 o.k.?
5 volt power supply defective.
Replace board D3.
18 volt power supply defective.
LED VCC18 on board D3 o.k. ?
Circuit diagram J1048 – 11
Connection cable D3.X8 o.k.?
Check the power supply for the pressure sensors!
17 volt power supply defective
LED VCC17 on board D3 o.k. ?
Replace board D3.
Valve and card terminal fuse defective?
Short circuit?
Fuse F4 on board D3?
Circuit diagram J1048 – 13
Check valve (coupling circuit).
Check card terminal.
Reference for temperature sensors defective.
Replace temperature sensors.
Replace board D3.