Text After Replacing Board D3 - Siemens MODULARIS Uro Plus Service Instructions Manual

Hide thumbs Also See for MODULARIS Uro Plus:
Table of Contents

Advertisement

Service software

Text after replacing board D3

Press the "h" key to display the following text (the software is installed).
--------------<
-------------------------------------------
n: select system name before shipping !
-------------------------------------------
software version: V_ _ _
select option:
==============
h: show this help screen
p: show pulse counters
...
...
...
o: change orientation of treatment table. Now:
(X) = Physician's place for
Press the "n" key to display the following text
=> N
select number to define system name:
1: SIEMENS
2: OEM-customer
Press 1 or 2 and follow the instructions.
you selected: SIEMENS
press 'C' to confirm, any other key to abort
SIEMENS
CAUTION
Siemens AG
Medical Solutions
xxxxxxx
Lithomodule
endourolog. therapy
This procedure cannot be undone.
SPL1-130.840.02
Rev. 03
>--------------
hardware:_ _ - _ _
------
:
\_/
(X) :
:
---------------
Page 15 of 18
04.05
CS PS 24
3 - 15
------
:
:
:
MODULARIS Uro Plus
3

Advertisement

Table of Contents
loading