High Voltage Connector - Siemens MODULARIS Uro Plus Service Instructions Manual

Hide thumbs Also See for MODULARIS Uro Plus:
Table of Contents

Advertisement

LITHOSTAR MODULARIS
S
Fig. 15

High voltage connector

Remove the covers of the basic system unit and the shock wave head.
Remove the shock wave head (refer to the corresponding section).
Remove the cover at the charging unit above the high-voltage connection by removing
the screws (Fig. 13).
Remove the three clamps (N/Fig. 13) on the high-voltage connection.
Push the high voltage cable (HK/Fig. 16) in the direction of the shock wave head through
the corrugated tubing until the high-voltage connector is accessible (Fig. 15).
Remove the screws (S/Fig. 15) on the high voltage connector.
Push the metal housing (G/Fig. 16) above the high-voltage plug toward the back.
Open the cover (D/Fig. 16) of the high-voltage connector.
Remove the connection clamps (M/Fig. 17) in the old high-voltage connector.
Open the cable clamps (KS/Fig. 17) of the high-voltage connector.
Connect the new high-voltage connector; reconnect and secure it in the reverse order of
the removal.
Push the high-voltage cable back in the direction of the charging unit (in the basic unit)
and connect it (Fig. 13).
Mount the shock wave head assembly (refer to the corresponding section).
Attach all covers (refer to the corresponding sections).
Siemens AG
Medical Solutions
G
G
Fig. 16
SPL1-130.840.02
Rev. 03
D
Fig. 17
Page 15 of 34
04.05
CS PS 24
4 - 15
M
KS
4
MODULARIS Uro Plus