Siemens MODULARIS Uro Plus Service Instructions Manual page 103

Hide thumbs Also See for MODULARIS Uro Plus:
Table of Contents

Advertisement

MUP Menu Ultrasound Device
Selection of the MODULARIS cross on the SONOLINE Prima
Ultrasound localization
Fig. 3
In ultrasound localization a holder in which an ultrasound probe can be fastened is fitted
to the shock wave head.
Only the following probes can be used:
Press the "F4" key.
Press the "5" key.
Select the MUP with the trackball.
Press the
Enter the password.
Select "Litho Mode no X-ray" with the trackball.
Press the
Activate "Litho Mode no X-ray" with the trackball.
Press the
Press the
Deselect the screen saver on the SONOLINE Adara (refer to the SONOLINE Adara
operating instructions).
When a
Siemens AG
Medical Solutions
3.5C40S convex array 3.5 MHz
3.5/5.0 SI sector probe
key.
key.
key.
key two times.
or
probe is selected, the following cross (Fig. 3) is displayed.
SPL1-130.840.02
Rev. 03
Page 3 of 8
04.05
CS PS 24
8 - 3
8
MODULARIS Uro Plus

Advertisement

Table of Contents
loading

Table of Contents