Adjusting The 0° Position - Siemens MODULARIS Uro Plus Service Instructions Manual

Hide thumbs Also See for MODULARIS Uro Plus:
Table of Contents

Advertisement

5 - 2
ok
requires
correction
Fig. 1
A
Fig. 3
Adjusting the 0° position
Remove the cover of the shock wave head (refer to chapter 4).
Couple the LITHOSTAR MODULARIS to the SIREMOBIL Iso-C as described in the
operating instructions. Do not couple the motor yet.
Manually perform the orbital movement (angulation range approx. ± 30°).
Set the SIREMOBIL Iso-C to orbital and angulation direction 0°.
Couple the motor.
Flip up the iso-center phantom.
Switch FL ON briefly.
The ball must be in the center of the cross (refer to Fig. 1).
If required, adjust the shock wave head as shown in Fig. 3 (A or B):
- A: Loosen the screws (1/Fig. 2) on both
sides.
-
Shift it with the screws (2/Fig. 2).
-
Retighten the screws (1/Fig. 2) on
both sides.
MODULARIS Uro Plus
acceptable
2/Fig.2
SPL1-130.840.02
Rev. 03
Iso-center with X-Ray
3
4
Fig. 2
B
4/Fig.2
- B: Loosen the screws (3/Fig. 2) on both
sides.
-
Shift it with the screws (4/Fig. 2).
-
Retighten the screws (3/Fig. 2) on
both sides.
Page 2 of 8
04.05
CS PS 24
1
2
1
5
Siemens AG
Medical Solutions