5 - 2
ok
requires
correction
Fig. 1
A
Fig. 3
Adjusting the 0° position
•
Remove the cover of the shock wave head (refer to chapter 4).
•
Couple the LITHOSTAR MODULARIS to the SIREMOBIL Iso-C as described in the
operating instructions. Do not couple the motor yet.
•
Manually perform the orbital movement (angulation range approx. ± 30°).
•
Set the SIREMOBIL Iso-C to orbital and angulation direction 0°.
•
Couple the motor.
•
Flip up the iso-center phantom.
•
Switch FL ON briefly.
The ball must be in the center of the cross (refer to Fig. 1).
•
If required, adjust the shock wave head as shown in Fig. 3 (A or B):
- A: Loosen the screws (1/Fig. 2) on both
sides.
-
Shift it with the screws (2/Fig. 2).
-
Retighten the screws (1/Fig. 2) on
both sides.
MODULARIS Uro Plus
acceptable
2/Fig.2
SPL1-130.840.02
Rev. 03
Iso-center with X-Ray
3
4
Fig. 2
B
4/Fig.2
- B: Loosen the screws (3/Fig. 2) on both
sides.
-
Shift it with the screws (4/Fig. 2).
-
Retighten the screws (3/Fig. 2) on
both sides.
Page 2 of 8
04.05
CS PS 24
1
2
1
5
Siemens AG
Medical Solutions