Siemens MODULARIS Uro Plus Service Instructions Manual page 95

Hide thumbs Also See for MODULARIS Uro Plus:
Table of Contents

Advertisement

Iso-center with ultrasound
3
2
1
Fig. 19
Checking image tilt
NOTICE
Check whether the isocenter phantom (1/Fig. 19) is filled with water; if not, remove screw
(2/Fig. 19) and fill in water with a syringe (3/Fig. 19).
There may not be any air bubbles. Turn in the screw again.
Lift up the isocenter phantom.
Fill the water bladder with water (approx. 100 - 120 ml, i.e. 5 - 6 syringes, each 20 ml)
(1/Fig. 20):
1.
Fill a syringe with distilled water. Remove any air that may be trapped in the
syringe.
2.
Insert the syringe needle through one of the valves (2/Fig. 20) and empty the
syringe contents in the water bladder. The valve has a self-closing membrane.
3.
Repeat steps 1 and 2 until the water bladder is filled with approx.
100 - 120 ml water.
4.
After filling, there may not be any air bubbles in the water bladder. If there are any
air bubbles, remove them using the syringe.
Spread contact gel on the isocenter phantom.
Secure the water bladder (5/Fig. 1) with the holder (6/Fig. 1) to the isocenter phantom.
Spread contact gel on the ultrasound probe.
Use the knob
Move the probe forward (3/Fig. 1) until the isocenter crosshair is approx. in the center of
the SONOLINE Prima screen.
Siemens AG
Medical Solutions
The isocenter must be checked with every probe type and after
every replacement of a probe.
to set a value of "60" (D/Fig. 14).
SPL1-130.840.02
Rev. 03
04.05
2
Fig. 20
Page 13 of 16
CS PS 24
6 - 13
1
MODULARIS Uro Plus