Angulation Drive (C-Arm Drive) - Siemens MODULARIS Uro Plus Service Instructions Manual

Hide thumbs Also See for MODULARIS Uro Plus:
Table of Contents

Advertisement

LITHOSTAR MODULARIS

Angulation drive (C-arm drive)

Open the cover of the angulation drive.
Disconnect the motor cable coming from the unit at the clamp (K/Fig. 26).
Desolder the cables at the potentiometer (P/Fig. 26); when doing this, record the color
coding of the cables to the potentiometer connections to ensure correct reconnection.
Remove the old angulation drive completely.
Install the new angulation drive. Do not tighten the mounting screws yet (5/5a/Fig. 26).
Adjust the C-arm to 0°.
Couple the LITHOSTAR MODULARIS to the SIREMOBIL Iso -C.
- The pin (6/Fig. 27) should go easily into the guide; if not, rotate the motor via the control
panel of the LITHOSTAR MODULARIS.
- The screws can still be loosened (5/5a/Fig. 26). Follow the correct sequence when
tightening (5 up, 5 down, 5a up, 5a down/ Fig. 26).
Set the potentiometer to 500 ± 10 ohms or 5.0 ± 0.1 volts (center position).
The potentiometer setting can also be checked with the Service PV (Chapter 3).
Tighten the toothed belt.
Adjust the C-arm according to the description on pages 3 - 10.
Reattach the motor cover.