Iwaki Cooling Pump; Removal; Installation - Siemens MODULARIS Uro Plus Service Instructions Manual

Hide thumbs Also See for MODULARIS Uro Plus:
Table of Contents

Advertisement

LITHOSTAR MODULARIS

IWAKI cooling pump

Removal

Switch the system off.
Open the system cover.
Remove the fasteners on the pump.
Remove the motor connections on the terminal block.
Loosen the screws to remove the pump.

Installation

Tighten the pump.
Connect the motor connections.
Connect the fasteners on the pump (The black-marked water hoses must be connected).
Fill the cooling circuit (refer to the section "Filling the cooling circuit").
Siemens AG
Medical Solutions
SPL1-130.840.02
Page 13 of 34
Rev. 03
04.05
CS PS 24
4 - 13
4
4
4
MODULARIS Uro Plus

Advertisement

Table of Contents
loading

Table of Contents